AK 98 Dialysis Machine - Attentions
B:159
589 Start air detector test
589 Start air detector test
Fill venous drip chamber.
Appears:
In the beginning of the priming phase when the venous drip
chamber has not yet been filled.
Machine actions:
None.
Operator action:
When the blood pump is running, fill the venous drip
chamber using the level adjustment knob. The level of
priming fluid should be well above the air detector head.
When the venous drip chamber is filled, the machine will
finish the functional check.
Mute time:
-