A:46
AK 98 Dialysis Machine - Machine Description
HCEN12745 Revision 10.2016
Program version 2.xx
Halt button
When this button is pressed, the power supply to the
machine is interrupted. As soon as the button is
released the power returns and the machine performs
a recovery.
If this does not work, make a complete reset. Use the
main switch to turn off the machine. Press and hold
the halt button until the main switch indicator is unlit.
Release the halt button. Switch on the machine and
wait until the main switch indicator lights up. Press the
On/Off
button to start the machine.
Battery charge indicator
This yellow lamp is lit when the mains cable is
connected to the mains supply and the main switch is
switched on. It indicates that the battery charge is
ongoing. The lamp is marked BACH.
Battery connect indicator
This green lamp is lit if battery back-up has been
installed. The lamp is marked BACO.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...