AK 98 Dialysis Machine - Attentions
B:91
516 Chemical fillup volume is too low
516 Chemical fillup volume is too low
Check pickup tube and disinfectant container.
Appears:
In the fill-up phase during a chemical disinfection program.
Machine actions:
The disinfection program stops.
Operator action:
1.
Check the pick-up tube and the disinfectant container.
2.
Press
Confirm
to start the fill-up phase again.
Mute time:
Permanent.