B:48
AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
206 Incorrect acidic concentrate
206 Incorrect acidic concentrate
Check set value and connected concentrate.
Appears:
When the selected concentrate does not match the
concentrate connected to the machine.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1.
Check that the concentrate container is filled.
2.
Check that there is no air in the concentrate tube.
3.
Check what concentrate is set in the dialysis machine.
a)
Press the
Fluid
button.
b)
Select the
Conc
tab.
4.
Check that the concentrate that is set is the same as
the concentrate connected.
5.
Check that the proper concentrate has been connected
correctly.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...