AK 98 Dialysis Machine - Alarms
B:39
122 Restarted after power failure
122 Restarted after power failure
To continue press Confirm.
Appears:
When the machine has recovered from power failure.
Machine actions:
None.
Operator action:
1.
Press
Confirm
.
2.
Check power supply, mains cable and backup battery.