AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
542 Fluid tube sensor test
542 Fluid tube sensor test
To start fluid tube sensor test, remove fluid
tubes from safety couplings.
Appears:
When a dialysis fluid tube sensor test is stopped during a
disinfection or rinse program.
Machine actions:
None.
Operator action:
Remove and attach the dialysis fluid tubes from the safety
couplings to restart the test.
Mute time:
2 minutes.