AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
550 Functional check is stopped
550 Functional check is stopped
To continue functional check, connect
concentrate connectors to stand by port.
Appears:
During functional check if a chemical disinfection program
has previously been performed and the concentrate
connectors (or disinfectant connector) are not properly
connected to the machine.
Machine actions:
The functional check stops.
Operator action:
- Check that the concentrate connectors are properly
connected to the stand-by ports.
- Check that the disinfectant connector is properly
connected to the parking port.
It is not possible to connect the concentrates yet.
Mute time:
2 minutes.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...