AK 98 Dialysis Machine - Haemodialysis - Single needle treatment
A:107
5)
Attach the venous blood line. Follow the
steps in Section 4.2.5
attach the end of the venous blood line to
the dialyzer.
6)
Attach the venous expansion chamber.
Place the venous expansion chamber
upright in the holder. Make sure that the
expansion chamber clicks into its correct
position.
The edge inside the holder should fit into
the groove of the expansion chamber.
7)
Connect the end of the venous blood line
to the expansion chamber blood line.
8)
Connect the other blood line from the
expansion chamber to the dialyzer.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...