A:84
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
10)
Connect the arterial blood line with the
saline or priming solution bag. Check that
the connections are tight.
11)
Attach the arterial pressure transducer
protector of the arterial blood line to the red
arterial pressure transducer connector.
Press to pierce the membrane and then
turn to secure the connection.
12)
Close all relevant clamps.
CAUTION!
Check that all relevant
clamps are closed to prevent
undetected blood loss.
4.2.5
Attach the venous blood line
The venous blood line has blue connectors.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...