B:52
AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
210 Low battery power
210 Low battery power
The machine will turn off within one minute.
Appears:
During power failure when the battery power is too low.
Machine actions:
●
The machine will automatically be switched off.
●
Power failure sound will appear.
Operator action:
Consider to end the treatment manually.
The battery may need to be replaced.