AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
562 Improper auto start setting
562 Improper auto start setting
Ready time is too close to actual time. Adjust
ready time.
Appears:
When the machine is switched off and the time for when the
heat disinfection program or rinse program is set to be
finished is too close to present time.
Machine actions:
The machine does not switch off until the time setting has
been changed.
Operator action:
Change the finish time for the rinse program or heat
disinfection program.
Mute time:
2 minutes.