HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions
B:117
543 Fluid tube sensor test
543 Fluid tube sensor test
Remove tubes from safety coupling. To
interrupt test press Confirm.
Appears:
When the dialysis machine cannot restart the dialysis fluid
tube sensor test.
Machine actions:
None.
Operator action:
To continue the activated disinfection or rinse program:
1.
Check that the dialysis fluid tubes are properly attached
to the safety couplings.
2.
Press
Confirm
. The dialysis fluid tube sensor test is
bypassed.
As an alternative, consult an authorised technician.
Mute time:
2 minutes.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...