HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - DIASCAN
A:137
2)
Select the
Setup
tab.
3)
Press
Interval
.
4)
Select
60 minutes
,
30 minutes
or
Single
.
5)
Press
OK
to start the measurement.
Results
The result of the measuring is displayed in the clearance measurement readout field.
The clearance measurement readout field shows:
●
latest measured clearance value.
●
time of latest measuring.
●
current/estimated end of treatment Kt.
●
the
Diascan
button (opens the Diascan menu).
9.4
Check Kt/V
The dialysis machine can calculate Kt/V as a single measurement or at intervals.
If Kt/V measurement is required, distribution volume has to be set before the
measurement check takes place. The distribution volume is patient-related (based on
the patient's dry weight) and has to be properly estimated and set by the operator in
order to obtain a correct Kt/V value.
Distribution volume is the urea distribution volume in litres (water in the body),
estimated for the individual patient, and based on the patient’s dry weight. The
estimation is based on certain formulae used for this purpose, chosen by the
prescribing physician responsible for the patient. The distribution volume has to be set
by the machine operator in order for a correct Kt/V value to be obtained.
NOTE!
To get a reliable Kt/V value, it is very important that the dialysis machine calculates
with an accurate distribution volume for the patient. Therefore fill in the patient
information with care in Distribution volume.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...