AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
529 Diascan clearance measurements paused during
isolated UF
529 Diascan clearance measurements paused
during isolated UF
Clearance measurement restarts when isolated
UF is complete.
Appears:
During an attempt to activate the clearance measurement
function or a clearance measurement check (single or
continuous) during ongoing isolated UF phase.
Machine actions:
The clearance measurement will automatically be paused
during the Isolated UF phase. The clearance measurement
starts automatically when the isolated UF phase is
complete.
Operator action:
The attention disappears after 30 seconds.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...