B:154
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
581 No BPM cuff is attached
581 No BPM cuff is attached
Check BPM connectors. To clear attention press
Confirm.
Appears:
When the blood pressure monitor detects that the cuff is not
attached to the machine.
Machine actions:
After 2 minutes there is a buzzer sound.
Operator action:
●
Check for kinks or leakage from the cuff and the cuff
hose.
●
Check that the cuff hose is properly connected to the
machine.
Mute time:
Permanent.