AK 98 Dialysis Machine - Attentions
B:151
578 New Ultrafilter required
578 New Ultrafilter required
{0} disinfections since last change.
Appears:
When the ultrafilter needs to be replaced. Too many
disinfections performed since last change.
The number of disinfections between the last change and
the appearance of the attention can be preset.
Machine actions:
None.
Operator action:
Replace the ultrafilter and confirm that the ultrafilter has
been replaced.
Mute time:
Permanent.