HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions
B:149
576 Negative UF rate
576 Negative UF rate
Check patient for weight gain. Check for
obstructions.
Appears:
When there is a backfiltration of 500 mL to the patient or a
backfiltration rate >50 mL/min for 5 minutes.
Machine actions:
None.
Operator action:
1.
Check if there are obstructions in the outlet tube (the
tube where the used dialysis fluid flows out from the
dialysis machine).
2.
Remove any obstructions and press
Confirm
.
3.
Check that the UF rate is correct. If not, contact an
authorised technician.
Mute time:
2 minutes.
WARNING!
If no action is taken for this attention, the patient may get
incorrect weight loss during the treatment.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...