A:154
AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
5)
Select
Rinse
.
6)
Press
OK
.
7)
Select day.
8)
Set the time when the program should
finish.
9)
Press
OK
.
10)
To schedule more days in the week: repeat
steps 7-9.
10.5.3
To turn off rinse program for a certain day
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Tools
tab.
3)
Press
Auto schedule
.
4)
Select day.
5)
Press
Off
.
6)
Press
OK
.
10.6
Machine storage with chemical disinfectant
10.6.1
Fill the dialysis machine with chemical disinfectant
If the dialysis machine is not to be used for more than 7 days it is recommended to fill
it with a chemical disinfectant before storage. It is recommended to use peracetic acid
as disinfectant during storage.
This function has to be preset by an authorised service technician.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...