AK 98 Dialysis Machine - Attentions
B:133
559 Heparin pump is overloaded
559 Heparin pump is overloaded
Check heparin line for obstruction.
Appears:
When the pressure in the syringe is too high during priming.
Machine actions:
The heparin pump stops.
Operator action:
Check the syringe. Check for kinks on the thin line
connected to the syringe. See Section 4.2.6
on page A:89 in Operators handbook.
Mute time:
2 minutes.