HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:97
3)
Press the settings one by one and set the
desired treatment parameters.
4)
If a bolus dose of heparin solution is to be
administered at treatment start, press
Heparin bolus volume
and set a bolus
volume in mL suitable for the patient.
4.2.12
Connect the patient
Before you begin
WARNING!
Check that the blood lines and the patient accesses (needles or catheters) are
tightly connected. If not, the patient could suffer from blood loss or air embolism.
WARNING!
To minimise the risk of damage to the patient’s blood cells, check that there are no
kinks in the blood lines and that the needles are of sufficient size.
The machine may not be able to detect all flow and pressure conditions, which in
combination with narrowed passages in the blood circuit, may lead to damage of the
blood cells.
WARNING!
Using a central venous catheter with potential negative pressure can cause air
embolism if the arterial access is disconnected and if infusion is connected before
the blood pump.
WARNING!
When using a catheter, keep in mind that a negative pressure may occur from the
catheter; check that the arterial line is clamped before connecting any infusion
before the blood pump.
WARNING!
The operator must take proper precaution in order to prevent coagulation in the
extracorporeal circuit.
Coagulation may lead to:
●
inadequate delivery of dialysis.
●
risks associated with movement of blood clots to the patient.
●
disabling of the air detector function if blood clots aggregate in the drip chamber.
WARNING!
To avoid patient blood loss and blood hemolysis, it is important to follow the
manufacturer’s instructions for use of the dialyzer and the blood lines. When
attaching the blood lines always ensure that all connections are properly secured
and that no part of the blood lines have been kinked. If the heparin pump is not to be
used, especially ensure that the cap at the end of the thin line used for heparin
solution, is properly closed.
CAUTION!
Select
Connect Patient
before connecting the blood lines to the access. This
allows the machine to detect blood and to start the treatment.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...