AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
219 UF rate stopped due to high venous pressure
219 UF rate stopped due to high venous
pressure
Check blood access, blood lines and venous
pressure transducer protector for
obstruction.
Appears:
When the dialysis machine cannot achieve the set UF rate
since the patient’s venous pressure is too high.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1.
Check the patient’s blood access, the venous pressure
transducer and check for clots in the dialyzer. Take
appropriate actions when you find an issue.
2.
Make sure that the venous pressure is decreased by at
least 50 mmHg. The alarm will automatically disappear
when the actual venous pressure has been decreased.
3.
If it is not possible to decrease the actual venous
pressure, wait (maximum 5 minutes) until the dialysis
fluid enters the dialyzer again.
4.
If the alarm reappears, consider ending the treatment.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...