B:78
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
503 Battery failure
503 Battery failure
Blood pump will not be able to run in case of
power failure. To continue, change battery or
press Confirm.
Appears:
●
When the backup battery does not pass functional
check.
●
After treatment if the backup battery test fails.
If the functional check fails, the alarm sound will be delayed
for 1 minute.
Machine actions:
None.
Operator action:
Press
Confirm
.
Mute time:
Permanent.