HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:103
9)
Hold the arterial blood line just before the
pump segment. Turn the blood pump rotor
anticlockwise at the same time as you pull
the blood line out.
WARNING!
If any of the pressure
transducer protectors on the
blood lines contains blood or
appears to be damaged, the
dialysis machine may be
contaminated with blood. An
authorised service technician
must replace and clean the
components in the machine.
10)
Open the air detector door. Pinch the
venous drip chamber and pull it out.
11)
Disconnect and remove the arterial and
venous blood lines.
12)
Only for single needle treatment
:
Remove the venous expansion chamber
together with the blood line set.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...