AK 98 Dialysis Machine - Alarms
B:57
214 Pv and Qd settings in conflict
214 Pv and Qd settings in conflict
Increase Qd to 500 mL/min.
Appears:
When the dialysis machine cannot achieve the set UF rate
since there is a conflict between the patient’s venous
pressure and the set dialysis fluid flow rate.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
1.
If possible, increase dialysis fluid flow rate to at least
500 ml/min:
- Press the
Fluid
button.
- Select the
Fluid flow
tab.
- Press
Dialysis fluid flow
and adjust the flow
rate.
2.
If you cannot increase the dialysis fluid flow rate,
decrease the venous pressure and wait, maximum 5
minutes, until the dialysis fluid enters the dialyzer again.
3.
If the alarm goes off again, consider ending the
treatment.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...