HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - DIASCAN
A:139
2)
Select the
History
tab.
3)
Press
History
. The view in
Diascan
History
will be different if distribution
volume has not been added.
9.6
Set a Kt/V target value
It is possible to set the desired Kt/V target for the patient to have achieved at the end
of the treatment. This value is called
Kt/V target
.
NOTE!
Before setting a Kt/V target, the dialysis machine needs to know the patient’s
distribution volume. See Section 9.4
on page A:137.
Procedure
1)
Press the
Diascan
button.
2)
Select the
Setup
tab.
3)
Press
Alarm settings
.
4)
Press
Kt/V target
.
5)
Enter a target value for Kt/V.
6)
Press
OK
.
7)
Press
Activate
and close the menu.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...