Dr
ä
ger Medizintechnik GmbH
Service Mode Savina Version 2.0
Test step 24 is used to display the following calibration data of the O
2
sensors (S3.1, S3.2):
Up to SW 1.01 -> not assigned.
As of SW 1.10 -> Test step 25 is used to test the LC display, the LEDs, and the 7-segment
displays. The following tests can be called up with
”
Test target
”
:
−
”
LED
”
: All LEDs can be switched on and off (LEDs for indication of voltage supply are not
triggered, they come on depending on voltage supply and charge state of the battery).
−
”
LCD
”
: The display is switched to
”
dark
”
for 5 seconds, then to
”
bright
”
for 5 seconds.
−
”
7-segment
”
: 7-segment displays, including decimal point, can be switched on and off.
"Cal.const"
Calibration
constant
"Offset voltage"
Electronics offset voltage
"Pressure"
Pressure at which calibration was carried out
"Temperature"
Temperature at which calibration was carried