Error List Savina Version 2.0
41.0070
Reference voltage of the O2 sensors is
faulty.
−
Operating v15 V prese
"Repair Procedure
Measuring the
supply voltages
")?
−
Operating v15 V for O2
sensors on O2 Diaphragm PCB
present (test point X4/14(+) and X4/17
(AGND) on the O2 Diaphragm PCB,
see also chapter "Schematics and
Diagrams
O2 DiaphragmPCB
−
Replace the O2 Diaphragm PCB.
41.0100
Monitoring of lower limit of operating
v5 V is faulty, detected by
SiSW.
−
See error code 410020.
41.0101
Monitoring of upper limit of operating
v5 V is faulty, detected by
SiSW.
−
See error code 410021.