Dr
ä
ger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
•
Install brand-new EEPROM.
•
Assemble the unit.
•
Switch on the unit.
A device malfunction is generated.
•
Adjust settings (as written down), time, and date on the unit.
Setting the clock will generate a new device ID (see also "
Error when updating from SW 1.00 to
1.10
" under chapter "Repair Information/Error Events").
•
Switch unit off and on again.
The unit should not generate any device malfunction. Settings recently made should be available.
The operating hours should have the same status as before replacement of the EEPROM. Should
O2 alarms be generated, you need to calibrate O2 sensor 2 manually (see also "
Instructions for
Use/Operating Instructions
"). Previously detected options must be activated as described below.