3.2.2
Earth leakage current test
Normal condition (N.C.):
I
≤
500
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
1000
µ
A
In the following steps the test is repeated, but
with the power plug turned over. This
condition can be created internally with some
types of test devices.
Normal condition (N.C.):
I
≤
500
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
1000
µ
A
Tester
Savina
Mains voltage
L/N
N/L
PE
PS
Earth leakage current test
Earth leakage current test
P
P
P
P