Error List Savina Version 2.0
63.0013
O2 sensor no. 1: Temperature
measured value is outside of
specification (not used for the time
being).
−
63.0014
O2 sensor no. 1: Compensation with
default pressure.
−
Check ambient pressure sensors in
Service Mode, test 27, and adjust if
necessary.
63.0015
O2 sensor no. 1: Compensation with
default temperature.
−
Check device temperature in Service
Mode, test 16, and expose device to
ambient temperature of 5 to 40
necessary.
63.0020
O2 sensor no. 2: INFO: Sensor not
fitted.
−
Install sensor.
63.0021
O2 sensor no. 2: INFO: Sensor is
faulty.
−
Replace sensor.
63.0022
O2 sensor no. 2: INFO: Sensor is not
operational yet.
−
Allow sensor to warm up.
63.0030
O2 sensor no. 1: INFO: Sensor not
fitted.
−
Install sensor.
63.0031
O2 sensor no. 1: INFO: Sensor is
faulty.
−
Replace sensor.