Dr
ä
ger Medizintechnik GmbH
Service Mode Savina Version 2.0
Test step 11 is used to test the emergency air valves V8/V9. The valves can be switched on
This test step is also used to display the expiratory pressure "Pawexp" (S4.1) and the inspiratory
pressure "Pawinsp" (S4.2).
V8/V9 off -> pressure reading approx. 94 mbar
V8/V9 on -> pressure reading maximum 5 mbar
A venting noise is clearly audible when the valves are switched to "on".
Test step 12 is used to test the calibration valves (V6.1, V6.2) of the airway pressure sensors
(S4.1, S4.2). The valves can be switched on or off. In addition, the inspiratory flow "Flowinsp", the
inspiratory pressure "Pawinsp", and the expiratory pressure "Pawexp" are displayed. The following
approximate values are displayed:
Connect (short-circuit) the inspiratory socket and the expiratory socket with a ventilation
hose and connect an external pressure measuring device to perform this test step.
Connect (short-circuit) the inspiratory socket and the expiratory socket with a ventilation
hose to perform this test step.