Service Mode Savina Version 2.0
Test step 21 is used to display the measured value of the internal A/D converter of the front
processor.
VCC+24V_MESS
0655 ... 0737 3,20 V ... 3,60 V Measurement signal of the +24
voltage.
VOP+15V_MESS
0573 ... 0655 2,80 V ... 3,20 V Measurement signal of the +15
voltage for the control electronics of the
bypass and expiratory valves.
VOP-15V_MESS
0573 ... 0655 2,80 V ... 3,20 V see
VOP+15V_MESS
SNGD_M
0002 ... 0010 0,01 V ... 0,05 V Connected ground on Control P
V_GOLD_CAP
0921 ... 1003 4,50 V ... 4,90 V (charged). Voltage of Gold-Cap
capacitor. Voltage is through-connected
via contacts of ON/OFF switch. The
capacitor is located on the Control PCB.
I_PIEZO_MESS
0000 ... 0008 0,00 V ... 0,04 V Measurement signal of alarm g
V_TEMP_INT
0184 ... 0246 0,90 V ... 1,20 V (at ambient temperature). Meas
signal for internal device temperature,
measured at Control PCB.
V_SYNC_PIEZO
0000 ... 1023 0,00 V ... 5,00 V (runs cyclically, due to silencing
Measurement signal for intermittent
tone of alarm generator.