4.2.2
Testing the keys
Call up test steps 1 and 2 in sequence (as of
SW 1.10)
Check the functioning of all keys.
Test value:
All keys should function properly.
4.2.3
Nebulizer flow test
Test Step 13
: call up this test step.
Connect a flowmeter tube with an oxygen
scale of up to 16 L/min to the nebulizer
connection. Switch on the nebulizer valve.
Test value:
Nebulizer flow = 8 L/min to 14 L/min.
4.2.4
Comparing measurement functions of the
flow sensors
Test Step 14
: call up this test step.
Connect (short-circuit) the inspiratory and the
expiratory sockets using a ventilation hose.
Set the "PRESSURE BYPASS" value to
40 mbar.
Test value:
The measured inspiratory and expiratory flows
must not differ by more than 13%.
4.2.5
Offset voltages of the airway pressure
sensors
Call up test
19
19
.
Test values:
"V_PAW_INSP" = 730 mV
±
100 mV
"V_PAW_EXP" = 730 mV
±
100 mV