Dr
ä
ger Medizintechnik GmbH
Software Download Savina Version 2.0
This description applies with effect from download software 9.1.1.
•
Connect Savina and PC (laptop) to mains power supply.
•
Connect PC and Savina to RS232 extension and to RS232 adapter.
•
Switch on the PC.
The optional
”
BIPAP
”
and
”
Autoflow
”
functions will be lost when updating the software
from SW 1.00 to SW 1.10 or from SW 1.01 to SW 1.10. They must be reactivat
entering the release number (release code). See also
”
Error when updating from SW
1.00 to 1.10
”
or
”
SW 1.01 to 1.10 update procedure
”
in chapter
”
Repair Information/Error
Events
”
.
The software cannot be loaded with a normal RS232 cable. Pins 1 and 9 must be
connected to a 10-k
Ω
resistor to enable the Control PCB to detect the SERVICE-Q
signal. Use only the specified cables and adapters.