Dr
ä
ger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
5.
The input value "Paw (SiSW)" shows an error.
1.
After a cold start and 15 s afterwards (only alarm).
2.
During standby and 15 s afterwards (only alarm).
1.
Generation of event "Paw stenosis 2" (warning "!!! Airway pressure high").
2.
Transmission of event "Paw stenosis 2" to ventilation control in order to relieve the
open the electrical safety valve.
3.
Detection 5: Alarm only.
1.
When airway pressure reaches 5 mbar (relief successful), but at the earliest after 50
2.
Five seconds after occurrence of event cause.
3.
Detection 5: The input value "Paw (SiSW)" does no longer show an error.
4.
If a measured value becomes invalid, the detected condition will not be restored as a re