![Dräger Savina Service Manual Download Page 28](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836028.webp)
28
A
B
C
4.1.4
Testing the external d.c. voltage supply
4.1.4.1
Battery o peration s imulation
“Connect
Charge tester (Evita), complete
" to
d.c. voltage input of Savina (use the two lower
sockets).
Using "
DC voltage source, 3 -18 V
" apply a
voltage of 14 V to the charge tes ter.
After 6 seconds, at the latest, the LEDs on the
front of Savina show the following operating
states:
Mains power supply: green
External battery: green
Internal battery:
Green: Internal rechargeable batteries are
fully charged.
Yellow: Internal rechargeable batteries are
being charged.
Reduce the voltage from the DC source from
14 V to 11 V.
The LEDs on the front of Savina show the
following operating states:
Mains power supply: green
External battery: yellow
Internal battery:
Green: Internal rechargeable batteries are
fully charged.
Yellow: Internal rechargeable batteries are
being charged.
Notice: Maximum test duration = 1 minute!
I
0
Right side terminal:
(+) red
Left side terminal:
(-) black
F
F