Dr
ä
ger Medizintechnik GmbH
Service Mode Savina Version 2.0
Test step 0 is used to read out the error log (error list). A list of error messages appears. 5 error
messages are visible at a time. Additional error messages can be viewed by scrolling. The error
messages have the following structure:
"Send Logbook" transmits the error log to a PC (for settings in the terminal program, see
also"
Entering the PC Service Mode
"). "Delete Logbook" deletes the error log.
For assignment of error numbers, see chapter "
Error List
".
Test steps 1 and 2 are used to test the keys on the front panel. The name of the key appears, e.g.
"Nebulizer", and "Not press." (not pressed). "Pressed" is displayed as soon as the key is pressed.
This method can be used to test all keys.
—
see
Test Step 1
—
First occurrence
Last occurrence
Error number
Number of errors