Error List Savina Version 2.0
40.0011
DC internal is not being charged.
−
May occur at extremely high ambient
temperatures. Expose device to lower
ambient temperatures.
−
See also error code "
40.0010
−
Install new internal battery.
40.0012
DC internal interrupted.
−
Check fuse for internal battery.
−
Check cable connection from internal
battery to power pack.
−
Check cable connection positive
terminal of internal battery 1 to
negative terminal of internal battery 2.
−
See also error code "
40.0010
40.0013
Internal direct voltage is too low.
−
See also error code "
40.0010
40.1010
AD converter in power pack is faulty.
−
Replace power pack.
−
See also error code "
40.0010
40.1020
Power pack is faulty.
−
Replace power pack.
−
See also error code "
40.0010