Dr
ä
ger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
1.
Expiratory valve faulty.
1.
The measured end-expiratory pressure PEEP(t) differs from the PEEP
target value
by more than
-5 mbar (PEEP(t) < PEEP
target value
-5 mbar).
2.
The measured end-expiratory pressure PEEP(t) differs from the PEEP
target value
by more than
+5 mbar (PEEP(t) > PEEP
target value
+5 mbar) and there was no "air trapping" during PEEP
measurement.
1.
After a cold start and 15 s afterwards.
2.
During standby and 15 s afterwards.
1.
Generation of event "PEEP malfunction" (warning "!!! PEEP inop.").