Error List Savina Version 2.0
40.1110
Communication (2) between power
pack and Control PCB is faulty.
−
Could be due to a log error after fast
change of status. Power Savina from
mains supply.
−
Replace power pack.
40.1120
Communication (3) between power
pack and Control PCB is faulty.
−
Could be due to a log error after fast
change of status. Power Savina from
mains supply.
−
Replace power pack.
41.0000
Operating v15 V is too low.
−
Measure the operating voltage
(see "Repair Procedure
Measuring the
supply voltages
").
41.0001
Operating v15 V is too high.
−
Measure the operating voltage
(see "Repair Procedure
Measuring the
supply voltages
").