C
Dr
ä
ger Med izintechnik GmbH
Repair Information/Error Events Sav ina Version 2.0
If the unit cannot be switched of at the ON/OFF switch, proceed as follows:
1.
Set ON/OFF switch to position "OFF".
2.
Press "standby" key.
1.
Set ON/OFF switch to position "
2.
Removing fuse of internal batter
3.
Disconnect external d.c. supply.
4.
Unplug power plug from socket-
5.
Reconnect power supplies.
6.
Re-install fuse.
This fault may be caused by the software which does not detect the position of the
ON/OFF switch. The fault should no longer occur with software version 2.n.