Service Mode Savina Version 2.0
Test step 20 is used to display the measured value of the internal A/D converter of the front
processor.
V_PAW_INSP
0123 ... 0184 0,60 V ... 0,90 V Measurement signal of the insp
pressure sensor S4.2.
V_PAW_EXP
0123 ... 0184 0,60 V ... 0,90 V Measurement signal of the exp
pressure sensor S4.1.
V_FLOW
0004 ... 0012 0,02 V ... 0,06 V (with flow exp. = 0). Flow signal o
expiratory flow sensor S2.
V_FLOW_INSP
0061 ... 0123 0,30 V ... 0,60 V (with flow insp. = 0). Flow signa
inspiratory flow sensor S1.
V_AKKU_MESS
0681 ... 0953 3,33 V ... 4,66 V Measurement signal of interna
voltage.
VCC+15V_MESS
0573 ... 0655 2,80 V ... 3,20 V Measurement signal of the +15
voltage.
VCC-15V_MESS
0573 ... 0655 2,80 V ... 3,20 V Measurement signal of the -15
voltage.
VCC+5V_MESS
0501 ... 0542 2,45 V ... 2,65 V Measurement signal of the +15
voltage.