4.15
Testing the ventilation
4.15.1
Testing ventilation and O2 mixing in IPPV
mode
Set the following values:
Operating mode = IPPV, AutoFlow OFF,
Pmax OFF,
Vt = variable,
Tinsp.= 5 s,
f = 6 1/min,
O2 = 40 vol%,
PEEP = 5 mbar,
Paw alarm limit = max
Test set-up:
Savina
Test lung
O2 analyzer
Pressure gauge —30 to 120 mbar