Dr
ä
ger Med izintechnik GmbH
Repair Information/Error Events Sav ina Version 2.0
Check fuse for internal battery.
Replace fuse.
Check voltage of internal battery in Service
Mode
Test Step 4
.
Is fuse OK?
Voltage within
tolerance?
Yes
No
Yes
No
Power pack faulty.
Voltage less
than 10 V?
Yes
No
Replace battery.
Remove internal battery (see Maintenance Procedure, chapter "
Replacing the
internal rechargeable batteries
"), recharge with external charger and re-install.
Or connect external battery (see
"Starting aid" for charging of internal battery
).
Observe charging specifications of battery.