![Dräger Savina Service Manual Download Page 121](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836121.webp)
F
o
r
i
n
t
e
r
n
a
l
u
s
e
o
n
l
y
.
C
o
p
y
r
i
g
h
t
r
e
s
e
r
v
e
d
.
R
5
6
6
4
9
0
0
F
L
1
.
f
m
2
7
.
0
8
.
0
1
A
B
C
Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
11.0301
Data retention (EEPROM), checksum
error: Data structure in EEPROM
(Magic).
−
Complete new data input in EEPROM
MP (see error codes 11.0300 and
11.0302 through 11.0306).
−
Replace EEPROM FP.
11.0302
Data retention (EEPROM), checksum
error: Software options (not used for
the time being).
−
11.0303
Data retention (EEPROM), checksum
error: DeviceID (not used for the time
being).
−
11.0304
Data retention (EEPROM), checksum
error: Settings:
−
Resetting of language version under
"Config".
−
Replace EEPROM FP.
11.0305
Data retention (EEPROM), checksum
error: Safety parameters.
−
Switch on Savina again.
−
Replace EEPROM FP.
11.0306
Data retention (EEPROM), checksum
error: Power pack variant (not used for
the time being).
−