Dr
ä
ger Medizintechnik GmbH
Software Download Savina Version 2.0
•
Connect Savina and PC to the mains power supply.
•
Connect the PC to the RS232 extension.
•
Connect the Savina to the RS232 adapter.
•
Interconnect the RS232 extension and the RS232 adapter.
•
Install the "Bootstrap Download" service program on the PC.
Under Windows NT the screensaver must be deactivated, otherwise the software of the
medical equipment could be damaged during downloading.
The software cannot be loaded with a normal RS232 cable. Pins 1 and 9 must be
connected to a 10-k
Ω
resistor to enable the Control PCB to detect the SERVICE-Q
signal. Use only the specified cables and adapters.