Error List Savina Version 2.0
42.0120
Micro-filter blocked (not used for the
time being).
−
42.0130
Micro-filter obstructed (detected by
SiSW).
−
Replace the micro-filter.
42.0200
Error with speed adaptation detected
by SW.
−
Report error to DS L
ü
beck.
42.0210
Speed limited due to high ambient
temperatures.
−
Expose device to lower ambient
temperatures.
43.0000
Internal temperature of device above
60
°
C.
−
Expose device to lower ambient
temperatures.
−
Check whether power pack fan and
device fan are functioning properly
43.0010
Device internal temperature sensor:
Short-circuit.
−