Dr
ä
ger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
•
Write down new device ID and serial number and send them to "User Help Desk" of Intensive
Care department stating the following:
For the attention of Mr. Matthias Hurlebaus
Note: Replacement of EEPROM on Control PCB
Please send me the release codes for the following options:
"Enter options available before replacement of EEPROM here".
Address:
Fax: (++49) 451/882 4089
Phone: (++49) 451/882 4646
E-mail: [email protected]
•
If release codes are available, activate "BIPAP" and "Autoflow" options by entering the
codes in the fourth page of the "Config" menu.
•
Switch unit off and on again.
•
Check whether options available before replacement of EEPROM here exist.
•
Perform the safety check and functional check as per the Test Certificate.