Service Mode Savina Version 2.0
VCC+5V_S_MESS
0491 ... 0532 2.40 V ... 2.60 V Measurement signal of the +5V s
voltage of the absolute pressure
sensors S6, S7. The voltage is
generated on the O2 Valve PCB from
the +15 V.
V_ABS_PRESS_1
0798 ... 0880 3.90 V ... 4.30 V (at normal atmospheric pressu
Measurement signal of the absolute
pressure sensor S7. The sensor is
located on the O2 Valve PCB.
V_ABS_PRESS_2
0798 ... 0880 3.90 V ... 4.30 V (at normal atmospheric pressu
Measurement signal of the absolute
pressure sensor S6. The sensor is
located on the O2 Valve PCB.
V_FIO2_2
0102 ... 0205 0.50 V ... 1.00 V (if system is free from O2).
Measurement signal of O2 sensor 2.
The measurement signal is processed
on the O2 Diaphragm PCB.