Dr
ä
ger Med izintechnik GmbH
Repair Information/Error Events Sav ina Version 2.0
"BIPAP" and "Autoflow" options are lost when updating software from SW 1.00 to SW 1.10. They
must be reactivated by entering the release number (release code). Activate options as
1.
Carry out download using SW 1.10.
2.
Generate device identification number (device ID).
In order to be able to generate the device ID you need to completely adjust the real-time
clock (tt.mm.jj / hr.min). Then switch the unit off and on again, and read out the device I
from page four of the "Config." menu.
3.
Write down new device ID and serial number and send them to "User Help Desk" of I
Care department stating the following:
For the attention of Mr. Matthias Hurlebaus
Note: Software update from SW 1.00 to SW 1.10
Please send me the release codes for the "BIPAP" and "Autoflow" options:
Address:
Fax: (++49) 451/882 4089
Phone: (++49) 451/882 4646
E-mail: [email protected]
4.
If release codes are available, activate "BIPAP" and "Autoflow" options by entering
release codes in the fourth page of the "Config" menu.